Pulse Tube Cryocooler (Part 4) - Valve Controlled

Follow
Hyperspace Pirate

This is the 4th part in my video series on designing/building/testing a pulsetube cryocooler in an attempt to create low enough temperatures to liquify nitrogen (196C or 77K). In this video, i'll be investigating an entirely different type of pulse tube that uses two valves one connected to a compressor and another vented to the atmosphere to create the pressure oscillations in the pulse tube. This is known as a "GiffordMcMahon" (GM)cycle, and is more commonly used in a refrigeration cycle with a moving displacer, but can also be found in use with pulse tubes.

The GM cycle is less efficient than using a piston (stirlingtype) to generate pressure oscillations, but has the advantage that it's cheaper and easier to build, since it can be coupled with a variety of different compressors and matching acoustic impedance is not neccesary. GMtype cycles typically operate at a COP of under 5% of Carnot, whereas advanced stirlingtype coolers can approach 30% of Carnot. Despite this, GM cycles are the primary type used in terrestrial applications, such as cryocoolers in labs and production
facilities
where modest quantities (under 10L per day) of LN2 are required. Despite the increased energy consumption, this configuration is typically cheaper to install/maintain than stirling cycle coolers, which are usually reserved for aerospace applications, like being mounted onboard sattelites or missiles to cool optics.

For the valve assembly, i used a pair of solenoid valves activated by MOSFETs which could have their timing controlled by an arduino with a front panel display, allowing me to know the exact timing I dialed in, which was very convenient for repeatability. Real GM cycle coolers use a rotary valve connected to a motor inside the pressurized assembly, which is more reliable than solenoid valves.

The entire assembly runs at around 0.5 hz. Typical commercial units range between 12 hz, in contrast to stirling cycle units, which can operate anywhere from 20 hz to 100 hz. However, pressure ratios are typically much higher on a GM unit, ranging from 23 with a baseline pressure typically from 1030 atmospheres, whereas pressure ratios on stirling units are usually in the neighborhood of 1.11.3. Most GM units will have a high pressure side around 200400 psi and a low pressure side from 50200 psi, depending on the cold head and compressor being used.

The lowest temperature i achieved running on compressed air was 83C, but I didn't consider this a valid result, because I wasn't able to repeat it. The lowest repeatable temperatures were in the low 70's, basically matching the performance of my previous stirlingtype pulse tube. Considering that the power input required was close to 1 kW, compared to the previous design's power of ~120W max., it's clear that this setup is far less efficient.

One of the major sources of inefficiency is the fact that my lowpressure side is simply connecting to the atmosphere, rather than being connected back to the compressor in a closed loop. This is because my current compressor isn't hermetically sealed, and not meant to operate with input pressures over 1 atm. With my high side pressure at 100 psi, theoretically, the most efficient low side pressure would be somewhere between 4060 psi.

In the next part of this series, I'll be generating hydrogen and using it as a working fluid instead of air, with a closedloop system running off a hermetically sealed fridge compressor. This will dramatically increase internal heat transfer, making the whole system more powerful and efficient, and should bring me much closer to my goal of 196C.

Part I:

   • Pulse Tube Cryocooler  Part 1  

Part II:

   • Pulse Tube Cryocooler  Part 2 (75C)  

Part III:

   • Pulse Tube Cryocooler (Part 3)  