STERIS InnoWave Ultrasonic Ultra+ In Service Video 1080

Follow
STERIS Healthcare United States

Visit the STERIS University course to earn a Certificate of Completion or CE Credit here: https://university.steris.com/course/...

Cette vidéo propose des soustitres français, en sélectionnant dans le menu Paramètres YouTube.

This video shows the proper use of the InnoWave Ultra Plus Sonic Irrigator Including:
• 0:00 Introduction
• 0:54 Using the Control Panel
• 2:57 Preparing Chemistry for Use
• 3:41 Loading Standard Ultrasonic Baskets
• 4:42 Selecting and Starting a Cycle
• 5:29 Aborting a Cycle
• 5:47 Unloading Instrument Baskets
• 6:10 Maintenance Tasks

Follow your facility's policies and industry standards regarding the use of ultrasonic cleaners and sonic irrigators as part of your cleaning process. Refer to the product’s User Manual for complete use details.

Some of the key information included in the inservice video includes:

Using the Control Panel
Use the start menu to begin an automated cycle sequence when it’s time to start the wash cycle. The wash menu lets you select the correct cycle from the list of preprogrammed and customizable cycles. Use the chemical menu to prime the chemical pump. Use the clock menu to set the time and date. Simply select the parameter you wish to change and use the arrow keys, change the values, then press the back arrow in the top left corner to return to the main menu.

To download cycle information, select the Data Logging menu, then insert a USB drive to initiate the download. You can choose to either download all cycle logs or just the new cycles since the last download.

Preparing Chemistry for Use
You will need to connect chemistry before using your ultrasonic. At the initial install, and every time chemistry is being changed, the lines need to be primed to ensure proper dosing and prevent cycle errors.

Loading an Ultrasonic Basket
The standard basket can hold up to 11 lbs. of lumened and nonlumened surgical instruments. To begin, place instruments in the basket, ensuring they do not overlap. Once the basket is properly loaded, gently place the basket inside the tank aligning it with the guiderails. The handles will lock the basket in place.

Selecting a Cycle
The InnoWave Ultra+ has three preprogrammed cycles and three spots for custom cycles. Before selecting and starting a cycle, complete all precycle checks. This includes verifying enough chemistry is present for the next wash, inspecting and cleaning the tank filter, and confirming the wash tank is free from debris.

Aborting a Cycle
Occasionally, you may need to stop a cycle before it is complete. Press the cycle pause button to pause the cycle. You will now see the 'Cycle Paused' screen. Select 'Abort' to cancel the cycle. If the abort sequence is confirmed, the unit will automatically drain.

Maintenance Tasks
To keep your Sonic Irrigator in excellent condition, clean and maintain it regularly. Remember to disconnect power from the unit and wear proper PPE when performing maintenance. Follow all facility procedures and refer to the user manual for full maintenance information.

Daily, spray and wipe down all internal surfaces above the waterline using a disinfectant and a clean cloth or paper towel, including the inside of the lid and around the seals. Wipe all external surfaces, including around the sides of the doors. Remove any trace of cleaner from internal and external surfaces with a clean wet cloth and then towel dry.

To learn more about InnoWave™ Ultra Plus Sonic Irrigator or request a quote, please visit: https://bit.ly/3KroL4E